The syringe printhead utilizes a stepper motor to compress a syringe plunger. The rate of extrusion, time of extrusion, and retraction volume can be control...
Fri, 27 Sep, 2019 at 2:51 PM
The maximum extrusion rate is 10 uL/sec.
Fri, 27 Sep, 2019 at 2:51 PM
The temperature range of the syringe printhead is from several degrees above room temperature to 65 °C.
Fri, 27 Sep, 2019 at 2:52 PM
The syringe printhead does not have any cooling functionality, you must use the temperature controlled printhead. It is not recommended to use the syringe p...
Fri, 27 Sep, 2019 at 2:52 PM
The syringe printhead has been optimized for the specified inner and outer syringe diameters. Due to the precision needed for bioprinting, it is not recomme...
Fri, 27 Sep, 2019 at 2:52 PM
You can use glass and metal syringes if they are the correct diameter. The movements of the syringe head were optimized for plastic syringes so it is recomm...
Fri, 27 Sep, 2019 at 2:53 PM
See the G-code list below to manually code the Syringe Printhead and customize functionality. Commands Description G1 Xnnn Ynnn Znnn Ennn Fnnn When use...
Fri, 27 Sep, 2019 at 4:46 PM
For BIO X's Syringe Printhead user manual, please feel free to refer to the attached user manual below for your reference. For further assistance and q...